Dead-Weight Procedure
Primary-Standard 4-Point Workflow
Dead-weight calibration is the primary-standard method and the default workflow for load cells below 10,000 lb rated capacity. The sequence below performs 4-point calibration (0%, 25%, 50%, 100% of rated capacity) with a final recovery check at 0% to confirm zero return.
Step 1. Prepare Reference Weights and Verify Traceability
Confirm the reference weights are within calibration (NIST Class F certificate current; recalibrated within the past 12 months). Verify total stack equals rated capacity within Class F tolerance.
Step 2. Warm Up and Stabilize the Indicator
Apply power to the indicator 30 minutes before calibration begins. Ambient temperature should be stable within ±2°C for the duration of the procedure.
Step 3. Zero the Indicator With Platform Empty
Confirm the load platform, fixture, or test rig is empty. Press Zero / Tare on the indicator and wait for reading stability (±1 display digit for ≥10 seconds).
Step 4. Apply Weights in 25% Increments and Record Readings
Place 25% rated load, wait 10 seconds for stability, record reading. Repeat at 50%, 75%, 100%. Calculate deviation at each point against expected reading (rated capacity × fraction).
Step 5. Verify Deviation Is Within Accuracy-Class Tolerance
For C3 (±0.02% OIML R60), deviation at each point must be within ±0.02% of rated output. For commercial class (±0.1%), tolerance is wider. If any point fails, investigate cell damage, cable issue, or indicator fault before adjusting.
Step 6. Remove Weights and Verify Zero Return
Unload fully. Reading should return to within ±0.01% of initial zero. If zero does not return, the cell has hysteresis beyond spec — schedule full replacement or return for NIST-traceable recalibration through Transcell calibration services.