How they work
All of the load, through one dome.
A load button is a compression-only load cell with the force applied to a domed button at the centre of its body. The dome is not decoration. It is the alignment system, and it is why the reading repeats in a fixture that is never quite square.
The dome decides where the contact is
Press a flat-topped load cell against a flat platen and the contact patch is decided by whatever flatness the fixture happens to have. Tilt the platen by a fraction of a degree (a shim under one corner, a ram with clearance in its guides) and the two faces touch along one edge. The resultant moves out to the rim, and a body machined to read axial compression is reading a bending moment as well.
A crowned button removes that variable. It touches a flat face at the top of the crown whatever the platen is doing within a small angle, so tilt migrates the contact a short distance across the crown instead of throwing it to the edge. The line of action stays near the axis and the gauges see the load they were trimmed for. The HC-LB sheets call it a central loading dome for self-aligning, repeatable force transfer.
Go deeper
Why size is the whole argument for this geometry
Every other compression geometry needs somewhere to put itself: a pancake wants a bolt circle, a column wants height and a threaded load path, a shear beam wants a cantilever length and a free end. A load button wants a flat patch to sit on and clearance for the dome.
The HC-LB puts 15,000 lbf into a body 3.23 in across and 1.77 in tall including the dome. At the top of the range 65,000 lbf (300 kN, thirty tonnes) sits inside 6.20 in of flange and 2.13 in of height. That is why they go into press tooling, jigs, end effectors and test fixtures, and why they are specified late, when the only volume left is a pocket in a plate.
Two families, two ways of writing the accuracy
The TC-LB sheet states nonlinearity, hysteresis and repeatability in absolute pounds-force per capacity, not as one percentage across the family: the 100 lbf model within ±0.15 lbf, the 1,000 lbf within ±2.5 lbf, the 10,000 lbf within ±25 lbf. Read your capacity's row; the error is not a constant fraction of rated.
The HC-LB is written the same way but resolves to one figure: nonlinearity ±7.5 lbf at 15,000, ±12.5 at 25,000, ±17.5 at 35,000, ±32.5 at 65,000: ±0.05% of rated every time, hysteresis the same, repeatability tighter at ±3 lbf on the 15,000. Creep is separate at ±0.05% of load over thirty minutes, and under a dwelling press it matters more than nonlinearity.
Sealing, and the number that tells you it has failed
The TC-LB body is 17-4PH stainless rated IP64: dust tight and splash protected, a test-lab rating rather than a washdown one. The HC-LB is all stainless, hermetically sealed to IP68, with a bonded foil strain gauge element: the version for a vessel foot or a pocket where condensation collects.
On a part this small the element sits close to the surface and the internal volume is tiny, so moisture across the bridge shows quickly as a wandering zero. Measure insulation resistance against the sheet: ≥500 MΩ at 50 V DC on the TC-LB, >5 GΩ on the HC-LB.